Glass Vial Container Closure Integrity: How the WILCO R48 MC P/V Tests Every Vial


A glass vial can look correctly sealed and still have a leak. For sterile pharmaceutical products, the ability of a container-closure system to maintain its integrity is fundamental to product quality and patient safety. Visual inspection has an important role, but it cannot establish whether a closure is leak-tight.


Container closure integrity testing (CCIT) uses an appropriate, validated method to assess package integrity under defined conditions. The WILCO R48 MC P/V available from GlobePack is a rotary, non-destructive leak-testing system designed to inspect every vial passing through its configured inspection path. This specific machine — serial number 080044 — is configured for 3 mL glass vials.


How the WILCO R48 MC P/V tests individual vials


The machine handles vials automatically through a defined testing sequence:


•  Infeed and transfer. A conveyor, infeed screw and entrance star wheel deliver vials to the rotary table.

•  Individual test stations. The rotary table contains 48 stations. Each vial is positioned within its own mechanically closed test bell.

•  Leak measurement. The selected test programme applies controlled differential-pressure or differential-vacuum conditions and evaluates the measured response against defined limits.

•  Automated separation. Following the test, vials are routed to the appropriate accepted, retest or reject outlet.


The P/V designation refers to the machine’s pressure and vacuum testing capabilities. These are selectable test methods within the programme setup; it would be incorrect to suggest that every vial automatically undergoes both methods.


The machine-specific operating manual also describes a station self-check using a test leak of known size. This provides a useful functional check of the measuring system, but it does not replace qualification of the equipment or validation of the test method for a customer’s product.


100% inspection — and what that actually means


The R48 MC P/V is designed for 100% vial-by-vial inspection: rather than drawing a sample, it tests each vial processed through the inspection system. This is an important capability for manufacturers seeking comprehensive in-line leak testing.


However, 100% inspection coverage does not mean every possible defect is guaranteed to be detected. The ability to identify a leak depends on the container, closure, contents, selected test method, test settings and validated acceptance criteria.


Glass Vial Container Closure Integrity: How the WILCO R48 MC P/V Tests Every Vial


Key specifications of GlobePack’s machine


•  Current configuration: 3 mL pharmaceutical glass vials (exact container and closure fit to be confirmed).

•  Inspection system: 48 rotary test stations; individual vial testing with automated separation.

•  Test technology: Selectable differential-pressure and differential-vacuum methods.

•  Detectable leakage: >10 microns, as stated in the OEM specification for this unit.

•  Quoted output: Up to 600 vials/minute, according to GlobePack’s listing; achievable speed depends on the selected programme and application.

•  Condition: Manufactured in 2009; listed as new and unused, in its original packing cases.


The OEM’s “>10 microns” specification is useful technical information, but should not be interpreted as a universally validated detection threshold for every vial or product. The pharmaceutical manufacturer must establish that the method, challenge conditions and acceptance limits are suitable for its specific container-closure system.


What EU GMP Annex 1 actually requires


EU GMP Annex 1 distinguishes between closure types. Under §8.22, glass ampoules, BFS units and small-volume containers (100 mL or less) closed by fusion are subject to 100% integrity testing using validated methods. For containers closed by other means, such as conventional stoppered and crimped glass vials, §8.23 requires samples to be tested using validated methods and a scientifically justified sampling plan.


Annex 1 §8.30 separately addresses individual inspection of all filled parenteral containers for extraneous contamination or other defects. This inspection is not interchangeable with CCIT. Section 8.25 also requires transport and shipping stresses to be considered when validating container closure integrity.


The WILCO R48’s 100% vial-testing capability may strengthen a manufacturer’s container-integrity control strategy, but it is a machine feature — not a blanket Annex 1 obligation for all stoppered glass vials, nor an automatic declaration of GMP compliance.


Suitability, qualification and recommissioning


As with any specialised pharmaceutical inspection system, the critical question is whether this particular machine is suitable for the intended vial and product. Before commissioning, manufacturers should assess container dimensions, closure design, product characteristics, test programme, performance at the required line speed, qualification and validation requirements, and site-specific data handling and controls.


This unit was manufactured in 2009 but is listed as unused. Following extended storage, its mechanical and control condition should be assessed before installation. The machine-specific manual calls for inspection by a WILCO technician before commissioning after storage exceeding 12 months.


Glass Vial Container Closure Integrity: How the WILCO R48 MC P/V Tests Every Vial


Request the WILCO R48 MC P/V specification


GlobePack can provide the available machine documentation, details of the 3 mL vial configuration and demonstration material, and discuss suitability for your product and production requirements. The demonstration video on our listing shows an identical machine for illustration; it is not a documented running trial of serial number 080044.


View the machine: WILCO R48 MC P/V – GlobePack listing


Contact GlobePack to request the specification, operating manual or demonstration video, or to discuss your vial and product.


Glass Vial Container Closure Integrity: How the WILCO R48 MC P/V Tests Every Vial


Technical references: WILCOMAT R48 MC P/V machine-specific English operation manual (serial 080044); GlobePack equipment listing; European Commission — EU GMP Annex 1 (§§8.22, 8.23, 8.25 and 8.30).


New Wilco R48MC Leak Detection System for Vials for Sale in Normanton, United Kingdom (globepackequipment.com)